HZBS-3 ASTM D93 Closed Cup Flash Point Apparatus
Huazheng HZBS-3 ASTM D93 Closed Cup Flash Point Apparatus HZBS-3 Closed Cup Flash Point Tester, color touch screen instead of keyboard operation. Used to determine the closed flash point value of petroleum products. Adopt advanced foreign technology, convenient and fast, open, fuzzy control integrated...
HZBS-3 ASTM D93 Closed Cup Flash Point Apparatus
The HZBS-3 Closed Cup Flash Point Tester features a color touch screen interface that replaces traditional keyboard operation. It is designed to determine the closed cup flash point of petroleum products.
Incorporating advanced foreign technology, the instrument offers convenient and fast operation through its integrated open architecture, and modular structure. It complies with ISO 2719, Chinese national standards, ASTM D93, and other relevant specifications.
Main technical features and functions
1.The new high speed digital signal processor is adopted to work reliably and accurately.
2.Test, open cover, ignition, alarm, cooling, printing, the whole test process is completed automatically.
3.This instrument uses electric ignition;
4.Automatic measurement of atmospheric pressure and automatic correction of test results.
5.Using the newly developed high power high frequency switching power heating technology, the heating efficiency is high, and the adaptive PID control algorithm is adopted to automatically adjust the heating temperature curve.
6.Temperature overvalue automatically stops detection and alarm.
7.Thermal - sensitive micro-printer, makes printing more beautiful and faster, with offline printing function.
8.Historical test records with time markers are stored up to 50.
9.640x480 color touch screen, easy to operate.
10.Using full-screen touch screen keys, intuitive and convenient operation.
11.Built-in multiple execution standards to choose from.
Main Technical Index
| Item | Specification |
| Temperature measuring range | -49.9 ℃ ~ 400.0 ℃ |
| Temperature control range | Room temperature ~ 250 ℃ |
| Repeatability | 0.029X (X = average of two successive tests) |
| Resolution | 0.1 ℃ |
| Precision | 0.5% |
| Temperature sensor | Platinum resistor (PT100) |
| Flash fire detection | K‑type thermocouple |
| Ambient temperature | 10 ℃ ~ 40 ℃ |
| Relative humidity | < 85% RH |
| Power supply | AC 230 V |
| Power consumption | < 500 W |
Precautions and Instrument Maintenance
The instrument should be used in a non-corrosive environment.
The oil cup should be cleaned when replacing the test sample.
Check that the thermocouple is free from oil contamination; otherwise, detection sensitivity may be affected.
Do not touch the ignition head, oil cup, or surrounding parts during the test to avoid burns.
For instruments using electric ignition, the ignition head requires regular replacement. Replace the ignition wire after every 1,000 ignitions (approximately 80–100 tests).








Hot Tags: hzbs-3 astm d93 closed cup flash point apparatus, China hzbs-3 astm d93 closed cup flash point apparatus manufacturers, suppliers, factory, flash point apparatus price, Flash Point Apparatus, pensky martens apparatus astm d93 , automatic cleveland open cup flash point tester, Cleveland Open Cup Flash Point Tester, closed cup flash point tester