HZ5372 2000A Primary Current Injection Tester
HZ5372 Primary Current Injection Tester Products Description 1 Input voltage: AC220V ±10% 2 Power: 5KVA 3 Output current: Series: AC0~1000A Parallel: AC 0~2000A stepless adjustable, 4 Output voltage: (>10V) 5 Current stability: 0.2% Basic Operations 1. Transformation Ratio Detection...
HZ5372 2000A Primary Current Injection Tester
Product Description
| Item | Specification |
| Input voltage | AC220V ±10% |
| Power | 5KVA |
| Output current | Series: AC0~1000A; Parallel: AC 0~2000A stepless adjustable |
| Output voltage | >10V |
| Current stability | 0.20% |
Basic Operations
1. Transformation Ratio Detection
Initial Setup: Before connecting the power cord, rotate the "Current Adjustment" knob counterclockwise to its zero position. Connect the current output terminals of the device to the primary side of the Current Transformer (CT) under test. Connect the device's ammeter to the secondary side of the CT.
Power On & Output:
Switch on the main power and turn on the instrument's power switch. Set the "Power Control" switch to the "ON" position; the power indicator should light up.
Set the "Selector Switch" to the required position (Series or Parallel).
Rotate the "Current Adjustment" knob clockwise slowly until the desired output current is reached.
Reading & Calculation: You can now perform the characteristic test on the device under test.
The primary ammeter displays the primary current value.
The secondary ammeter displays the secondary current value.
The transformation ratio of the CT under test can be calculated based on the measured primary and secondary current values.
Selector Switch Settings:
Series Position: Terminals L2 and L3 must be short-circuited. The output terminals are L1 and L4. Maximum output current is 1000A.
Parallel Position: Terminals L1 and L2 must be short-circuited, and terminals L3 and L4 must be short-circuited. The output terminals are (L1 and L3) or (L2 and L4). Maximum output current is 2000A.
2. Impulse Test
First, preset the required impulse current value using the "Current Adjustment" knob.
Press the "Impulse" switch. The device will power off momentarily.
Release the "Impulse" switch. The device will instantly output the preset impulse current, allowing you to perform the impulse test on the equipment under test.