HZKM-2 Engine Oil Anti-friction And Wear Testing Machine
HZKM-2 Engine Oil Anti-friction And Wear Testing Machine Products Description Purpose: To test the lubrication, anti-wear performance, and oil temperature changes of lubricating oils both in their base form and when blended with anti-wear additives. Test Conditions: Under 220V power, conventional oil...
HZKM-2 Engine Oil Anti-friction And Wear Testing Machine
Product Description
Purpose
The instrument is designed to evaluate the lubrication performance, anti-wear characteristics, and oil temperature behavior of lubricating oils. Testing can be conducted on base oils as well as oils blended with anti-wear additives, allowing for direct comparison of performance enhancements.
Test Conditions
The test is conducted under a 220 V power supply. Under specific applied pressure conditions, conventional oil will seize-indicating a failure of the lubricating film-whereas wear-resistant lubricant formulated with anti-wear additives maintains its lubricating properties and does not seize under the same conditions.
Optimized Operating Procedure
Prepare Shaft Sleeve:
Connect power & start motor.
Thoroughly polish shaft sleeve with ordinary lubricating oil using thumb pressure.
Wipe sleeve clean. Stop motor. [Purpose: Remove original oil film/nanomembrane for unbiased results.]
Setup Test Machine (Ordinary Oil Baseline):
Place steel ball in short lever socket.
Fill oil box with ordinary lubricating oil.
Install oil box, long lever, and short lever.
Run Baseline Test (Ordinary Oil):
Start motor.
Gradually increase load until seizure occurs.
Record torque wrench reading at seizure.
Monitor ammeter: Increasing current indicates rising power/oil consumption.
Note increasing noise.
Remove short lever. Observe ball wear (depth, area, roughness).
Reset & Prepare for Test Oil:
Reposition steel ball. Reinstall lever assembly.
Repolish shaft sleeve [to remove wear from baseline test]. Wipe clean.
Refill oil box with wear-resistant lubricating oil.
Run Test with Wear-Resistant Oil:
Gradually increase load on long lever.
Observe: Noise decreases significantly. Ammeter shows minimal current/power increase (indicating fuel/oil savings).
Note: Machine does not seize despite high load.
[Conclusion: Test oil demonstrates excellent anti-wear properties.]
Verify Nanomembrane & Conclude:
Remove oil box. Observe machine runs normally. [Confirms nanomembrane presence/effect.]
Stop motor.
Remove steel ball. Compare wear marks to baseline test results.
| Parameter | Specification |
|---|---|
| Model | HZKM-2 (Lever Type, Digital Display) |
| Instrument Type | Lever-type lubricating abrasion / wear resistance tester |
| Loading Method | Lever + balancing weights (progressive) |
| Display | Digital display |
| Power Supply | AC 220V |
| Test Principle | Rotating shaft sleeve vs. loaded steel ball; seizure load + current measurement |
| Test Oil Capacity | Via oil cup (2 pcs supplied) |
| Measurement Indicators | Ammeter (current change), torque wrench reading, steel ball grinding mark |
Products Cases

Company Profile

Certification

Related Products
Send Inquiry