Withstand Voltage Tester Operating Procedures
May 14, 2024
Operation test
1. Connect one end of the 0.7MΩ standard resistor to the ground wire of the withstand voltage instrument.
2. Turn on the power supply and set the instrument and alarm leakage current to 5mA.
3. Turn on the instrument and hit the other end of the standard resistor with a test rod. Adjust the voltage to within 3410V to 3590V. If the instrument alarms, it is determined that the instrument is in normal working condition. If the instrument alarms outside the range of 3410V to 3590V, the instrument is not working properly.
4. When the equipment function fails during the operation inspection and the operation inspection results cannot meet the specified requirements, the operator needs to re-test the products that have been tested since the last qualified operation inspection and send the instrument for repair.
Operation steps
The operator's seat and feet must be padded with rubber insulation pads. Only when the test light is off and there is no high voltage output can the tested model be connected or disassembled.
1. Calibrate the instrument before testing. Method: Under the leakage current of 5mA, use a 700KΩ ceramic resistor to bridge between the ground clamp and the high voltage test rod probe until the instrument alarms.
2. When connecting the model under test, make sure that the voltage meter is set to "0", and when the test light is off, clamp the instrument ground wire clamp to the heat sink of the machine under test, and press the power switch of the machine under test.
3. Set the instrument test conditions: A. Voltage: 3500V; B. Leakage current: 5mA; C. Test time: 4 seconds for assembly line production.
4. Place the test rod probe close to any AC input metal plug at the power cord head.
5. Press the start button to observe the test results. Within the set time, the over-leakage light is off, and the model under test is qualified.
6. If the model under test exceeds the set leakage current value, the instrument automatically cuts off the output voltage, and the buzzer alarms at the same time. The over-leakage light is on, and the model under test is unqualified. Press the reset button to clear the alarm sound. When testing again, press the start button again.